Đầu ra: 500VA (AC: 5kV/100mA)
Điện áp ra: 0.2kV~5.00kV AC,100mA 10V/step
Tần số ra: 50Hz, 60Hz (optional)
Sóng đầu ra: sóng sin, trở kháng thuần, trên 200V, biến dạng <5%
Giới hạn trên dòng dò: 0.10~100.00mA
Giới hạn dưới dòng dò: 0.10~100.00mA
Thang đo dòng đánh thủng: 0.100~3.499mA/3.50~100.00mA±2.5%
Thiết lập thời gian kiểm tra: 1.0~999.9S
Kiểm tra bù dây: 0~0.500mA
Giao tiếp: RS232(RS485 optional), Remote control(PLC)
Kích thước: W×H×D(mm): 450×150×440
Product Detail
Main features
* Latest international standard:Comply with CCC,IEC,EN,VDE,BS,UL,JIS standard requirements.
* Voltage ramp up/down,satisfy latest industrial standard requirements
* Breakdown current upper/lower limit setting,intelligent judgment,sound and light alarm
* AC 50/60HZ output
* Over voltage,over current protection,manual test wires compensation
* VFD display,simple operation
* Remote,RS232/485 interface
* Satisfy low voltage distribution industry latest requirements
Spec.
|
Model
|
AN9602X(F)
|
AN9605X(F)
|
AN9632X(F)
|
AN9671X(F)
|
|
AC Hipot Test
|
Rated Output
|
500VA
AC: 5kV/100mA
|
100VA(AC: 5kV/20mA)
|
|
|
Output Voltage Set and Range
|
0.2kV~5.00kV AC,100mA 10V/step
|
0.2kV~5.00kV AC, 10V/step
|
|
|
Output Voltage Adjustable Rate
|
±(2.5% of setting+10V)
|
|
|
Output Voltage Resolution/Accuracy
|
10V, ±(2.5% of setting+10V)
|
|
|
Output Frequency Range
|
50Hz, 60Hz (Optional)
|
|
|
Output Waveform
|
Sine waveform, pure resistance, above 200V, distortion <5%
|
|
|
Upper and Lower Limit Leakage Current Setting
|
AC: Upper: 0.10~100.00mA, Lower:0.10~100.00mA(0=disable), 0.01mA/step
|
AC: Upper: 0.10~20.00mA, Lower:0.10~20.00mA(0=disable), 0.01mA/step
|
|
|
Breakdown Current Measurement Range
|
0.100~3.499mA/3.50~100.00mA
|
0.100~3.499mA/3.50~20.00mA
|
|
|
Breakdown Current Accuracy
|
±(2.5% of reading+5 counts)
|
|
|
Ramp-up Time Setting Range/Resolution
|
1.0~999.9s (0=disable) 0.1s/step
|
|
|
Test Time Setting Range/Resolution
|
1.0~999.9s (LLL=continues) 0.1s/step
|
|
|
Test wire compensation
|
0~0.500mA, manual clear offset
|
|
|
Insulation Resistance Test
|
Output Voltage Setting and Measurement Range
|
|
100V ~1000V DC 5V/step
|
|
Output Voltage Resolution/Accuracy
|
|
1V , ±(3% of setting+5V)
|
|
DC Ripple Coefficient<span style="font-s
|